14
Installation
Remote Terminal Strip
For receiving start, control, and error signals from external devices
For sending start, control and error signals to external devices
Events Remote Connector
For test purposes or in some other cases, it can make sense to manually
enter these signals:
Sending control signals (Events) to external devices
Opening and closing contacts
Legend
1
Display
2
Events
Remote Con-
nector
Fig. 25
Remote terminal strip
1
2
Signal
Explanation
ANALOG:
OUT
Analog output signal. Either outputs the measured
system pressure or a control voltage for Pump B. The
output range can be set to values of max. 1, 2, 5 and
10 V.
ANALOG: IN
Analog input signal for
controlling the flow rate, e.g.:
1 V for 1 ml/min in the case of the 10 ml pump head
1 V for 5 ml/min in the case of the 50 ml pump
head.
START: OUT
Output is active for 500 ms when the pump starts.
START: IN
Activated by a voltage of 0 V against GND.
ERROR: OUT
Output is active until the Error condition has been
eliminated.
ERROR: IN
A voltage of 0 V against GND stops the pump.