15. Perform some temperature measurements
15.1. Perform a measurement of ambient temperature
➢
Connect a Pt100 temperature probe (available as option) on C1 connection on the top of the analyser.
➢
Turn on the instrument
Home screen is displayed with measurement menu highlighted.
➢
Press
OK
.
The analyser performs the auto-zero: remaining time and a progress bar display on the screen. However it is possible to
reach the temperature menu during the auto-zero.
➢
Go to
“Ambient Temp.”
with the keypad then press
OK
.
The temperature detected by the Pt100 probe is displayed on the analyser.
15.2. Perform a measurement of flue gas temperature
➢
Connect the flue gas probe on the analyser.
➢
Turn on the analyser.
Home screen is displayed with measurement menu highlighted.
➢
Press
OK
.
The analyser performs the gas auto-zero: remaining time and a progress bar display on the screen. However it is possible to
reach the temperature menu during the auto-zero.
➢
Go to
“Flue gas Temp.”
with the keypad then press
OK
.
The screen displays the flue gas temperature detected by the thermocouple probe of the flue gas probe.
15.3. Perform an ambient temperature measurement with the internal sensor
➢
Turn on the instrument
Home screen is displayed with measurement menu highlighted.
➢
Press
OK
.
The analyser performs the auto-zero: remaining time and a progress bar display on the screen. However it is possible to
reach the temperature menu during the auto-zero.
➢
Go to
“Internal Temp.”
with the keypad then press
OK
.
The screen displays the NTC temperature detected by the analyser.
15.4. Perform a Flow/Return temperature measurement
➢
Connect 2 thermocouples Velcro type (ref. : SKV 150) or lamella contact (ref. : SKCT) probes on the Kigaz.
➢
Turn on the instrument.
Home screen is displayed with measurement menu highlighted.
➢
Press
OK.
The analyser performs the auto-zero: remaining time and a progress bar display on the screen. However it is possible to
reach the temperature menu during the auto-zero.
➢
Go to
“Flow/Return Temp.”
with the keypad then press
OK
.
The screen displays the T1 temperature and the T2 temperature and the delta of temperature (ΔT) detected by the analyser.
15.5. Link the measurement to a customer
When the measurement has been performed, it is possible to link it to a customer registered in the analyser or to a customer
unregistered in the analyser that must be created.
➢
Press the
“Save”
function button.
The analyser displays the list of recorded customers.
➢
Select a customer with up and down arrows then press
OK
or create a customer (see page 23) then select it.
The analyser displays the list of available boilers.
➢
Select the boiler with up and down arrows then press
OK
or create a boiler (see page 24) then select it.
The analyser backs to the measurement screen.
15.6. Add the measurement to a group
The group function has to be activated (see page 18).
Once the measurement is linked to a customer and a boiler, it is possible to add it in a group in the instrument, or create a group:
The analyser proposes to add the measurement to a group.
➢
Go to
YES
and press
OK
.
➢
Go to the required group with the arrows and press
OK
, or create a group (see page 25).
32
Perform some temperature measurements