3
Performance Verification for Analog Analyzer
58
Keysight U8903B Service Guide
2
Connect the function generator outputs to the frequency counter input channel 1 and
2, and to the U8903B unbalanced input channel 1 and 2 as shown in
. Preset
the U8903B, function generator, and frequency counter.
Figure 3-8
Function generator to frequency counter and U8903B connection for wideband frequency
accuracy verification
3
Configure the function generator output channel as follows:
– Output load: High-Z load
– Output polarity: Normal
– Waveform function: Sine wave
– Amplitude unit = Vrms
– Amplitude offset = 0 V
4
Configure the frequency counter input channels as follows:
– If frequency counter model is 53220A, set SCPI language mode: 53132A
– Impedance: 1 M
Ω
– Low pass filter: ON (for frequency <100 kHz)
– Input coupling: DC
– Level at center of hysteresis window: 0 V
– Trigger level percentage of signal peak-to-peak range: 50%
– Reference timebase: Internal reference
– Measurement function: Channel 1 frequency
– Frequency measurement start arm mode: Immediate
– Frequency measurement stop arm mode: Timer
5
Configure the U8903B input channels to be tested as follows:
– Connection type: Unbalanced
– Input bandwidth: 90 kHz
Unbalanced input
channel 1
Function generator
Frequency counter
Unbalanced input
channel 2
U8903B audio
analyzer
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Страница 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 161: ...Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test 162 ...
Страница 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...