2
Performance Verification for Analog Generator
32
Keysight U8903B Service Guide
9
Set the Driving channel output state to ON and the Measured channel output state to
OFF.
10
On the U8903B analyzer, enable the output for both the Driving and Measured
channels.
11
Record the crosstalk measurement obtained from the Measured channel in the “
record for crosstalk verification
12
to
for the rest of the frequency, analyzer input range, and
generator output amplitude combinations to complete the verification of the U8903B
output channels to be tested.
13
to measure the generator crosstalk from channel 2 to channel
1 in which the Driving channel is channel 2 and the Measured channel is channel 1.
14
Compare the recorded crosstalk measurement values to the specification. If the test
fails, adjustment will need to be carried out accordingly.
Square wave rise time verification
In this test, the U8903B generates a square wave swept over a set of frequencies, and the
rise time is measured using a Keysight DSO8064A/8104A digital oscilloscope to ensure
that it is within the published specification.
1
Connect the U8903B Unbalanced output channel 1 to the digital oscilloscope channel
1 and Unbalanced output channel 2 to the digital oscilloscope channel 2 as shown in
. Preset both the U8903B and the digital oscilloscope.
Figure 2-5
U8903B to oscilloscope connections for square wave rise time verification
2
Configure the digital oscilloscope as follows:
– Trigger mode: Edge
Table 2-5
Square wave rise time specification
Rise time specification
Equipment used
<
2
μ
s
Keysight DSO8064A/8104A digital oscilloscope
Unbalanced output
channel 1
U8903B audio
analyzer
Digital
oscilloscope
Unbalanced output
channel 2
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Страница 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 161: ...Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test 162 ...
Страница 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...