![Keysight U8903B Скачать руководство пользователя страница 70](http://html.mh-extra.com/html/keysight/u8903b/u8903b_service-manual_1961458070.webp)
4
Performance Verification for Digital Generator and Digital Analyzer
70
Keysight U8903B Service Guide
7
Set the DSO8064A oscilloscope to measure the amplitude of Function 1 and record the
measurement as V
measure
.
8
Compute the level accuracy using the following equation.
Level Accuracy (dB) = 20
×
Log (V
measure
/ V
DUT
)
9
to
for the rest of the V
DUT
output level accuracy verification
to complete the verification of the
different voltages.
10
to
for the rest of the F
SR
as given in “
output level accuracy verification
to complete the verification of the
different sampling rates.
1.8
0.15
2.4
2.0
0.175
2.8
2.3
0.195
3.12
2,.5
0.21
3.36
2.8
0.235
3.76
3
0.25
4
3.3
0.275
4.4
3.5
0.295
4.72
3.8
0.32
5.12
4
0.335
5.36
4.3
0.36
5.76
4.5
0.375
6
4.8
0.4
6.4
5.1
0.425
6.8
Table 4-4
Channel 1 and channel 2 vertical scale and vertical scale range (continued)
V
DUT
(V)
CH1 and CH2 voltage per division (V)
Function 1 vertical scale range (V)
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Страница 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 161: ...Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test 162 ...
Страница 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...