Performance Verification for Digital Generator and Digital Analyzer
4
Keysight U8903B Service Guide
65
g
Set the Digital Output Sampling Rate based on the sampling rate (Hz) as given in
“
Test record for DSI master clock output frequency accuracy verification
.
h
Set the DSI Master Clock Multiplier based on the Master Clock Multiplier as given in
“
Test record for DSI master clock output frequency accuracy verification
.
i
Measure the Master Clock Rate twice and average the two measurements.
j
Record the average Master Clock Rate, F
measure
in the “
clock output frequency accuracy verification
.
k
Compute the master clock error using the following equation. MCLK is the master
clock rate calculated where MCLK = MUL
×
Sampling Rate (Hz).
Δ
F
error
ppm = [(F
measure
– MCLK) / MCLK]
×
1000000
6
Verification steps for word length at 16 bits:
a
Set the DSI audio resolution to 16 bits.
b
Set the DSI audio word length to 16 bits.
c
Set the DSI Master Clock multiplier to 128.
d
Set the digital output sampling rate to 48 kHz.
e
Set the DSI audio resolution based on the Word Length as given in “
DSI master clock output frequency accuracy verification
.
f
Set the DSI audio word length to 32 bits.
g
Set the Digital Output Sampling Rate based on the sampling rate (Hz) as given in
“
Test record for DSI master clock output frequency accuracy verification
.
h
Set the DSI Master Clock Multiplier based on the Master Clock Multiplier as given in
“
Test record for DSI master clock output frequency accuracy verification
.
i
Measure the Master Clock Rate twice and average the two measurements.
j
Record the average Master Clock Rate, F
measure
in the “
clock output frequency accuracy verification
.
k
Compute the master clock error using the following equation. MCLK is the master
clock rate calculated where MCLK = MUL
×
Sampling Rate (Hz).
Δ
F
error
ppm = [(F
measure
– MCLK) / MCLK]
×
1000000
7
Verification steps for word length at 20 bits:
a
Set the DSI audio resolution to 20 bits.
b
Set the DSI audio word length to 20 bits.
c
Set the DSI Master Clock multiplier to 160.
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Страница 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 161: ...Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test 162 ...
Страница 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...