Performance Verification for Analog Generator
2
Keysight U8903B Service Guide
25
Performance Verification
DC accuracy verification
This test verifies the accuracy of the DC signal generated by the U8903B over a range of
amplitudes using two digital multimeters to ensure that it is within the published
specification.
1
Connect the U8903B Unbalanced output channel 1 to the first digital multimeter and
Unbalanced output channel 2 to the second digital multimeter as shown in
.
The BNC-to-banana jack adapter is used for this connection. Preset both the U8903B
and digital multimeters.
Figure 2-1
U8903B to digital multimeter connections for DC accuracy verification
2
Configure the digital multimeter as follows:
– Measurement function: DC voltage
– Level filter: Enabled
– NPLC: 10
– Number readings: 1 (with auto sample event)
– Trigger arming: Hold (suspend)
– Trigger mode: Auto
– Range: Refer to “
Test record for DC accuracy verification
3
Configure the U8903B output channels to be tested as follows:
– Connection type: Unbalanced
– Output impedance: 600
Ω
Table 2-1
DC accuracy specification
DC accuracy specification
Equipment used
±
1.0%
2
×
Digital multimeter
Unbalanced output
channel 1
U8903B audio
analyzer
Digital multimeter 1
Digital multimeter 2
Unbalanced output
channel 2
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Страница 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 161: ...Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test 162 ...
Страница 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...