4
Performance Verification for Digital Generator and Digital Analyzer
80
Keysight U8903B Service Guide
12
Set the U8903B calibration digital output mode to NORM mode for AES/SPDIF.
13
Set the digital output sampling rate F
SR
as given in “
Test record for unbalanced input
sampling rate accuracy verification
.
14
Set the U8903B digital analyzer to measure the sampling rate and record the
measurement reading as SR
meas
.
15
Compute the sampling rate error using the following equation based on the previous
recorded F
SR_FC
value.
Δ
F
SRerror
ppm = [(SR
meas
– F
SR_FC
) / F
SR
]
×
1000000
16
for the rest of the F
SR
as given in “
unbalanced input sampling rate accuracy verification
to complete the
verification of the different output sampling rates.
17
to
with the digital unbalanced input impedance set to High
impedance (Hi-Z).