Performance Verification for Analog Analyzer
3
Keysight U8903B Service Guide
59
– Input coupling: AC
– Voltage detector type: RMS
– Trigger source: Immediate
– Input voltage range:1 V
– Measurement function2: AC Voltage
– Sample size: 1048576 for 20 Hz / 524288 for 50 Hz / 262144 for above 50 Hz
6
Set the function generator to produce a sine wave with amplitude of 1 Vrms and
frequency of f
DUT
as given in the “
Test record for wideband frequency accuracy
.
7
Set the U8903B input channels sample size as given in the “
frequency accuracy verification
.
8
Set the frequency counter gate time as given in the “
frequency accuracy verification
.
9
Enable the function generator output.
10
Using the
FETC? FUNC1,(@1)
SCPI command for the U8903B, record the frequency
measurement on the frequency counter, f
FC
and the U8903B channel 1, f
Meas
in the
Test record for wideband frequency accuracy verification
11
Compute the frequency accuracy of the U8903B using the following equation:
Wideband frequency accuracy = (f
Meas
– f
FC
/ f
FC
)
×
1000000
12
for the rest of the f
DUT
values to complete the verification of
the U8903B input channel 1 to be tested.
13
to complete the verification for the U8903B input channel 2.
14
Compare the recorded wideband frequency accuracy values to the specification. If the
test fails, adjustment will need to be carried out according
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Страница 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 161: ...Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test 162 ...
Страница 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...