Performance Verification for Bluetooth Card
5
Keysight U8903B Service Guide
93
A2DP
Bluetooth
audio and functional test points and limits
The A2DP Bluetooth audio and functional test uses a sine wave signal to check the
frequency accuracy, amplitude accuracy, flatness, and THD+N ratio for the analyzer and
generator.
shows the test points as well as the associated test limits for the maximum A2DP
audio bandwidth (48 kHz sample rate).
shows the test points as well as the associated test limits for the A2DP APTX
functional test.
Table 5-4
A2DP
Bluetooth
audio test points and limits for analyzer and generator
Frequency (Hz)
Amplitude (FFS)
Frequency limit (ppm)
Amplitude accuracy
limit (dB)
THD+N ratio limit (dB)
1000
1
±
10
-0.3 to 0.1
≤
-45
20
1
±
10
-0.8 to 0
≤
-45
50
1
±
10
-0.6 to 0
≤
-45
100
1
±
10
-0.3 to 0.1
≤
-45
500
1
±
10
-0.3 to 0.1
≤
-45
2000
1
±
10
-0.3 to 0.1
≤
-45
3000
1
±
10
-0.3 to 0.1
≤
-45
5000
1
±
10
-0.3 to 0.1
≤
-45
10000
1
±
10
-0.3 to 0.1
≤
-45
12000
1
±
10
-0.3 to 0.1
≤
-45
18000
1
±
10
-1 to 0
≤
-45
Table 5-5
A2DP APTX functional test points and limits for analyzer and generator
Frequency (Hz)
Amplitude (FFS)
Frequency limit (ppm)
Amplitude accuracy
limit (dB)
THD+N ratio limit (dB)
1000
1
±
130
-0.3 to 0.1
≤
-45
20
1
±
130
-0.8 to 0
≤
-45
18000
1
±
130
-1 to 0
≤
-18
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Страница 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 161: ...Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test 162 ...
Страница 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...