Command Reference
Tracking Generator Option Subsystem
Tracking Generator Amplitude Step
:SOURce:POWer:STEP[:INCRement] <value>
:SOURce:POWer:STEP[:INCRement]?
This command sets the amplitude step size of the tracking
generator output. This command is valid after tracking
generator is enabled.
Range:
0 to –10 dBm
Example:
:SOURce:POWer:STEP 1 dB
Key access:
Meas > Amptd Step
Tracking Generator Storing as Reference
:CALCulate:NTData:STORe
This command stores the current trace to Trace 4 as a
reference. This command is valid after tracking generator is
enabled.
Key access:
Meas > Normalize >Store Ref
Tracking Generator Normalization State
:CALCulate:NTData[:STATe] OFF|ON|0|1
:CALCulate:NTData[:STATe]?
Use this command to turn on/off tracking generator
normalization.
*RST:
Off
Example:
:CALCulate:NTData 1
Key access:
Meas > Normalize > Normalize On/Off