161
Command Reference
Channel Scanner Option Subsystem
Limit Type
:CHSCanner:DISPlay:LIMit:TYPE UPPer|LOWer
:CHSCanner:DISPlay:LIMit:TYPE?
This command toggles limit type between upper and lower.
Example:
:CHSCanner:DISPlay:LIMit:TYPE LOWer
Key access:
Shift > Limit > Limit Type
Display Sort Type
:CHSCanner:DISPlay:SORT ID|POWer
:CHSCanner:DISPlay:SORT?
This command toggles display sort type between by ID and
by power.
Example:
:CHSCanner:DISPlay:SORT POWer
Key access:
Shift > Disp > Sort By
Display Sort Order
:CHSCanner:DISPlay:SORT:ORDer DOWN|UP
:CHSCanner:DISPlay:SORT:ORDer?
This command toggles display sort order between down and
up.
Example:
:CHSCanner:DISPlay:SORT:ORDer UP
Key access:
Shift > Disp > Sort Order
Channel Scanner Logging Subsection
Logging State
:CHSCanner:LOG[:STATe] RUN|STOP
Use this command to start/stop long time logging of channel
scanner measurement result.
Example:
:CHSCanner:LOG RUN
Key access:
Meas > Logging Start