Command Reference
Sense Subsystem
80
Bandwidth Subsection
Resolution Bandwidth
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth. For numeric entries, all
RBW types choose the nearest (arithmetically, on a linear
scale, rounding up) available RBW to the value entered.
Range:
10 Hz to 3 MHz
Default Unit:
Hz
Example:
:BAND 1 kHz
Key access:
BW
> RBW
Auto Resolution Bandwidth State
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO
OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
This command turns on/off auto resolution bandwidth state.
*RST:
On
Example:
:BWID:AUTO On
Video Bandwidth
[:SENSe]:BANDwidth|BWIDth:VIDeo <freq>
[:SENSe]:BANDwidth|BWIDth:VIDeo?
Specifies the video bandwidth.
*RST:
3 MHz
Range:
1 Hz to 3 MHz. This range is dependent upon the setting of
[:SENSe]:BANDwidth|BWIDth[:RESolution]
.
Default Unit:
Hz
Example:
:BANDwidth:VIDeo 10 KHZ
Key access:
BW > VBW Auto Man