Command Reference
Reflection Measurement Option Subsystem
134
Reflection Measurement IIM State
[:SENSe]:CAT:IIM[:STATe] OFF|ON|0|1
[:SENSe]:CAT:IIM[:STATe]?
Use this command to turn on/off the IIM state for reflection
measurement.
Example:
:CAT:IIM ON
Key access:
Meas > IIM
Reflection Measurement Display Type
[:SENSe]:CAT:DISPlay:TYPE RL|VSWR|LIN
[:SENSe]:CAT:DISPlay:TYPE?
Toggles the reflection measurement display mode between
RL, VSWR and LIN. The RL stands for Return Loss, and LIN
stands for Reflection Coefficient. It’s only available for
Reflection Measurement mode, and Distance to Fault
measurement mode.
Range:
RL, VSWR, LIN
Example:
:CAT:DISPlay:TYPE VSWR
Key access:
Meas > Disp Mode
Set Reflection Measurement Cable Type
[:SENSe]:CAT:CABLe:TYPE S1000| S2000| S2500|CUST
[:SENSe]:CAT:CABLe:TYPE
Toggles the cable type between S1000, S2000, S2500 and
CUST. CUST stands for custom cable type. It’s only available
for Distance To Fault measurement.
Range:
S1000, S2000, S2500, CUST
Example:
:CAT:CABLe:TYPE S1000
Key access:
Meas > More 1 of 2 > Cable Specification > Cable Type