Command Reference
Power Meter Option Subsystem
208
Use this command to turn on/off limit state.
Example:
:CALCulate1:LIMit:STATe 1
Key access:
Shift > Limit > Limits
Set Limit Upper Value
:CALCulate1:LIMit:UPPer[:DATA] <value>
:CALCulate1:LIMit:UPPer[:DATA]?
This command defines the upper limit value.
Range:
–199.99 to +200 dBm
Example:
:CALCulate1:LIMit:UPPer:DATA 12
Key access:
Shift > Limit > Upper Limit
Set Limit Lower Value
:CALCulate1:LIMit:LOWer[:DATA] <value>
:CALCulate1:LIMit:LOWer[:DATA]?
This command defines the lower limit value.
Range:
–200 to +199.99 dBm
Example:
:CALCulate1:LIMit:LOWer:DATA -10
Key access:
Shift > Limit > Lower Limit
Set Limit Beep
:DISPlay[:WINDow]:DISPrange:BOTTom <value>
:DISPlay[:WINDow]:DISPrange:BOTTom?
Use this command to turn on/off beep warning for limit.
Example:
:CALCulate:LLINe:CONTrol:BEEP ON
Key access:
Shift > Limit > Limit Beep