101
Command Reference
Initiate Subsystem
Initiate Subsystem
The INITiate subsystem is used to control the initiation of
the spectrum analyzer sweep operation.
Continuous or Single Sweep
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Selects whether the sweep operation is continuously initiated
or not.
This command affects sweep if not in a measurement, and
affects trigger when in a measurement. This corresponds to
continuous sweep or single sweep operation when not in a
measurement, and continuous measurement or single
measurement operation when in a measurement.
When not in a measurement, this command does the
following:
•
When ON at the completion of each sweep cycle, the
sweep system
immediately initiates another sweep cycle.
•
When OFF, the sweep system remains in an “idle” state
until
CONTinuous
is
ON or
:INITiate[:IMMediate]
is
received. On
receiving the
:INITiate[:IMMediate]
command, it will go through a
single sweep cycle, and
then return to the “idle” state.
•
The query returns 1 or 0 into the output buffer. 1 is
returned when there is
continuous sweeping. 0 is
returned when there is only a single sweep.
When in a measurement, this command does the following:
•
When ON at the completion of each trigger cycle, the
trigger system
immediately initiates another trigger cycle.
•
When OFF, the trigger system remains in an “idle” state
until
CONTinuous
is
ON or
:INITiate[:IMMediate]
is
received. On
receiving the
:INITiate[:IMMediate]
command, it will go through a
single trigger cycle, and
then return to the “idle” state.