Sink Calibration Procedures
5
Jitter Tolerance Characterization Test
Purpose and Method
This procedure tests the jitter tolerance of a device for different sinusoidal
jitter frequencies. It is only available after configuring the product in
“Expert Mode”.
For each sinusoidal jitter frequency, this procedure tries to find the jitter
amplitude at which the target BER can be sustained. It starts by setting
the maximum jitter amplitude. Then, the value is decreased using either
binary or linear search algorithm. The range of jitter frequencies to be
tested can be defined in the “Swept SJ frequency” parameters section.
The result is a curve that shows the maximum jitter that the DUT can
tolerate as a function of the SJ frequency.
Because the limitations of the instruments and the setup have influence on
the result (for example, if the DUT would tolerate higher jitter amplitudes
than the generator could apply), the maximum possible (or applied) jitter is
also reported.
Connection Diagram
Figure 37
Connection Diagram for Receiver Tests
Содержание N5990A-155
Страница 1: ...User Guide Keysight N5990A 155 DisplayPort...
Страница 6: ...6 Keysight N5990A 155 DisplayPort User Guide Contents...
Страница 7: ...Keysight N5990A 155 DisplayPort User Guide 1 Introduction Overview 8 Document History 9...
Страница 10: ...10 Keysight N5990A 155 DisplayPort User Guide 1 Introduction...
Страница 79: ...Keysight N5990A 155 DisplayPort User Guide 5 Receiver Tests DisplayPort Receiver Tests 83...
Страница 104: ...104 Keysight N5990A 155 DisplayPort User Guide 5 Sink Calibration Procedures...
Страница 105: ...Keysight N5990A 155 Display Port User Guide 105...