Keysight N5990A-155 DisplayPort User Guide
43
Connection Diagrams
4
Random Jitter Calibration
Purpose and Method
This procedure calibrates the random jitter (RJ).
For this purpose, use the EZJIT+ application on the real-time scope. The
test automation starts by applying small RJ amplitudes and increases the
amplitude levels. For every set RJ value, the corresponding amplitude is
measured.
The Random Jitter Calibration procedure is available for each supported
data rate.
Connection Diagram
Figure 17
Connection Diagram for TP1 Calibration
Parameters in Expert Mode
• Calibration
•
Oscilloscope Bandwidth
•
Measurement Cycles
•
Memory Cycles
• Jitter
•
Use Explicit Clock
•
ISI Amplitude
•
Stop Jitter
•
Jitter Step Count
• Pattern