Keysight N5990A-155 DisplayPort User Guide
75
Connection Diagrams
4
Aggressor Amplitude Calibration
Purpose and Method
The Aggressor Amplitude Calibration procedure is available for each
supported data rate.
This calibration calibrates the differential output voltage of the aggressor
signal.
The test automation starts by applying a high differential voltage and
decreases it in each step. The corresponding aggressor signal amplitude is
measured on the scope.