66
Keysight N5990A-155 DisplayPort User Guide
4
Connection Diagrams
Eye Opening Calibration
Purpose and Method
This calibration calibrates the eye opening after the ISI emulator.
All the jitter impairments (RJ, ISI, SSC, SJFIXED, and SJSWEEP) are
applied to the data signal. Aggressor signals are not considered for this
calibration. The scope is configured to show the real-time eye.
For RBR and HBR, the test automation detects the crossing points of the
eye by using a histogram measurement and calculates the 50% crossing
point (middle of the eye). At this location, two histogram measurements
are used to measure the eye opening.
For HBR2, the test automation uses the location between 37.5% and
62.5% of the horizontal UI, where the eye opening is maximum, for the
measurement.
After the accomplishment of the aforementioned procedure, the generator
output voltage is decreased and the same measurement is repeated. This
allows an amplitude range to be calibrated. The eye opening is measured
and averaged over a number of cycles.
This calibration procedure is available for each supported data rate; for
TP2 and TP3; and for each Lane.
Содержание N5990A-155
Страница 1: ...User Guide Keysight N5990A 155 DisplayPort...
Страница 6: ...6 Keysight N5990A 155 DisplayPort User Guide Contents...
Страница 7: ...Keysight N5990A 155 DisplayPort User Guide 1 Introduction Overview 8 Document History 9...
Страница 10: ...10 Keysight N5990A 155 DisplayPort User Guide 1 Introduction...
Страница 79: ...Keysight N5990A 155 DisplayPort User Guide 5 Receiver Tests DisplayPort Receiver Tests 83...
Страница 104: ...104 Keysight N5990A 155 DisplayPort User Guide 5 Sink Calibration Procedures...
Страница 105: ...Keysight N5990A 155 Display Port User Guide 105...