Keysight N5990A-155 DisplayPort User Guide
3
Calibration and Test Procedures
Results
Run-Time Data Display
Most procedures generate data output. While the procedure is running,
the data is displayed in a temporary MS Excel worksheet or HTML page
(depending on the selected results viewer in the Station Selection), which
appear automatically for each individual procedure.
The MS-Excel worksheet or the HTML page that are displayed temporarily
during the procedure runs are thereby closed, once the specific procedure
is finished. As long as the N5990A Software is running, each worksheet or
page can be reopened with a double-click on the respective procedure.
However, the individual worksheets or pages are lost when the N5990A
main window is closed, unless you saved them.
Results Workbook
All calibration and test data worksheets can be saved any time in a
workbook by selecting “
File > Save Results as Workbook...
”. It is
recommended that this step be carried out at least at the end of each
N5990A run. If the calibration and test procedures are conducted several
times during the same instance of N5990A, the result worksheets are
combined in the workbook. If a test procedure is conducted without
performing calibration procedures prior to the same test run, only the test
results will be saved to the workbook.
NOTE
As a safety feature, all calibration and test results are saved by default to
the ValiFrame “Tmp” directory (refer to the
N5990A Getting Started
Guide
). The sub-folder “
Results/DisplayPort Station
” contains the Excel
files of the final results measured at each calibration and of the test
procedures. In addition to the calibration data worksheets, the
calibration data files are generated. These files are saved by default to
the N5990A calibrations folder. If these calibrations are run again, the
data files are overwritten. In order to save the calibration data files for
each configuration, the files must be copied from the directory:
C:\
ProgramData\BitifEye\ValiFrameK0\DisplayPort\Calibrations
and saved
manually in another folder before rerunning the calibrations.
Содержание N5990A-155
Страница 1: ...User Guide Keysight N5990A 155 DisplayPort...
Страница 6: ...6 Keysight N5990A 155 DisplayPort User Guide Contents...
Страница 7: ...Keysight N5990A 155 DisplayPort User Guide 1 Introduction Overview 8 Document History 9...
Страница 10: ...10 Keysight N5990A 155 DisplayPort User Guide 1 Introduction...
Страница 79: ...Keysight N5990A 155 DisplayPort User Guide 5 Receiver Tests DisplayPort Receiver Tests 83...
Страница 104: ...104 Keysight N5990A 155 DisplayPort User Guide 5 Sink Calibration Procedures...
Страница 105: ...Keysight N5990A 155 Display Port User Guide 105...