Keysight N5224B/25B Service Guide 4-7
Troubleshooting
Getting Started with Troubleshooting
Getting Started with Troubleshooting
Where you begin troubleshooting depends upon the symptoms of the failure.
Start by checking the basics as outlined in the following section. Also review
the flowchart in
. You should then be able to determine
where in the troubleshooting procedure to begin, to locate the failed assembly.
Check the Basics
A problem can often be solved by repeating the procedure you were following
when the problem occurred. Before calling Keysight Technologies or returning
the instrument for service, please perform the following checks:
1.
Is there power at the mains receptacle? If not, correct this situation and
proceed.
2.
Is the instrument turned on? Check to see if the front panel line switch
displays a light. This indicates the power supply is on. If the front panel
line switch is on but the power supply does not appear to be on, go to
“Power Up Troubleshooting” on page 4-9
3.
Is the Windows® operating system running? If not, refer to
System Recovery” on page 8-100
for instructions.
4.
If other equipment, cables, and connectors are being used with the
instrument, make sure they are clean, connected properly and operating
correctly.
5.
Review the procedure for the measurement being performed when the
problem appeared. Are all the settings correct? If not, correct them.
6.
If the instrument is not functioning as expected, return the unit to a known
state by pressing the
UTILITY > Main >
key.
7.
Is the measurement being performed, and the results that are expected,
within the specifications and capabilities of the instrument? Refer to the
embedded help in the analyzer for instrument specifications.
8.
If the problem is thought to be due to firmware, check to see if the
instrument has the latest firmware before starting the troubleshooting
procedure. Refer to
“Firmware Upgrades” on page 8-99
for instructions.
9.
If the necessary test equipment is available, perform the operator’s check
and system verification in
Chapter 3, “Tests and Adjustments.”
Troubleshooting Organization
Follow the flowgraph in
to help direct you to the correct section for
troubleshooting the analyzer.
Preset
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...