Keysight N5224B/25B Service Guide 7-19
Repair and Replacement Procedures
Removing and Replacing the A4–A17 Boards
Removing and Replacing the A4–A17 Boards
Tools Required
— T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
— T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
— 5/16 inch open-end torque wrench (torque to 10 in-lb or 1.13 N.m)
— 9 mm socket or open-end wrench (torque to 21 in-lb or 2.38 N.m)
— ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure. And, refer to
Assemblies, All Options, S/N Prefixes <6021”
Cables—All Options, S/N Prefixes <6021”
for this procedure.
1.
Disconnect the power cord.
2.
Remove the outer and inner covers. Refer to
.
3.
A5 and A10 source boards:
a.
A10 source board only: on the top side of the analyzer, disconnect
all visible semirigid cables from the A10, A12, and A13 boards (item
).
b.
A5 source board only: on the top side of the analyzer, disconnect all
visible semirigid cables from the A4, A7, and A8 boards.
c.
On the top side of the analyzer, remove two screws (item
), one at
each end of the board, from the source board to be removed.
d.
Lift the two extractors (item
), one at each end of the board.
Adjust the slack in the gray flexible cable to move it out of the way,
and lift the board out of the chassis.
4.
A7, A8, A12, A13 boards: on the bottom side of the analyzer, disconnect
cables (item
) from the cable brackets.
If you have a PNA with a serial number prefix <6021 and have not had a
2S7 or 4S7 upgrade, then you have A4, A14, A15, and A17 boards.
If you have a PNA with a serial number prefix
≥
6021 or have had a 2S7 or
4S7 upgrade, then you have an A15 direct digital synthesizer (DDS)
assembly instead of the A4, A14, A15, and A17 boards.
Refer to
“Top Assemblies and Cables, All Options, Serial Number Prefixes
“Top Assemblies and Cables, All Options, Serial
Any cables that are removed should be labeled for reinstallation later.
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...