Keysight N5224B/25B Service Guide 3-45
Tests and Adjustments
Adjustments
Step 5.
After pressing all of the cross hair calibration points a Digitizer Calibration
Tool window opens, press
Yes.
Step 6.
In the Tablet PC Settings window, press
OK.
Verification Procedure
Step 1.
Verify your instrument is powered up.
Step 2.
Press
System > Service > Display... > Display Setup > Calibrate
Touchscreen...
Step 3.
In the Table PC Setting window that opens, press
Calibrate...
Step 4.
Follow the instrument’s display prompts
.
10 MHz Frequency Reference Adjustment
The 10 MHz frequency adjustment is used to adjust the frequency accuracy of
the network analyzer’s 10 MHz frequency reference on the A14 frequency
reference board assembly.
Equipment Used for the 10 MHz Frequency Reference Adjustment
Procedure
1.
Connect the equipment as shown in
. Connect a GPIB cable
between the network analyzer and the frequency counter.
If your display fails the verification process, you can reset your display using the
Table PC Setting window. Refer to
“Calibration Procedure” on page 3-44
Equipment Type
Model or Part Number
Alternate Model or Part Number
Cable, BNC, 50W, 24 inch
8120-1839
Any
Frequency counter
53151A, Option 001
Any that will measure a signal at 10 MHz.
This adjustment typically adjusts to within
±
0.01 ppm.
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...