3-44 Keysight N5224B/25B Service Guide
Tests and Adjustments
Adjustments
3-
Adjustments
These adjustments are firmware-driven tests that are used to fine-tune your
analyzer.
If multiple adjustments are to be performed, perform them in the order listed.
—
“Touchscreen Adjustment and Verification” on page 3-44
—
“Synthesizer Bandwidth Adjustment” on page 3-46
—
“Source Adjustment” on page 3-46
—
“IF Gain Adjustment” on page 3-47
—
“Receiver Characterization” on page 3-47
—
“Receiver Adjustment” on page 3-48
—
“EE Default Adjustment” on page 3-50
—
“IF Response Adjustment (Available Only with Option S93093/4A and or
S93900/1/2/4/5/7A Installed)” on page 3-50
—
“LFE Receiver Adjustment – (Available with Option 205, 220, 405, 420, and
These adjustments are described on the following pages.
Touchscreen Adjustment and Verification
The touchscreen adjustment (calibration) is used to correct the touchscreen
accuracy when you replace your front panel, or your SSD. Refer to Calibration
Procedure.
The touchscreen verification can be run to verify the touchscreen has been
correctly calibrated. Refer to
“Verification Procedure” on page 3-45
.
Calibration Procedure
Step 1.
Verify your instrument is powered up.
Step 2.
Press
Instrument > Display > Display Setup > Calibrate Touchscreen...
Step 3.
In the Table PC Setting window that opens, press
Calibrate...
Step 4.
When the User Account Control opens, press
Yes.
For best results, use a touchscreen soft touch pen for the following
adjustments.
If your touchscreen is not behaving as expected, you can reset your display
using the Table PC Setting window and re-calibrate the touch settings
using the
below.
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...