3-46 Keysight N5224B/25B Service Guide
Tests and Adjustments
Adjustments
3-
Figure 3-15
Equipment Setup for the 10 MHz Frequency Reference Adjustment
2.
Press UTILITY
, then
, then
, then
click 10 MHz Freq Adjustment.
3.
Ensure the GPIB settings are correct.
4.
Follow the instructions and prompts as they are displayed.
Synthesizer Bandwidth Adjustment
This adjusts the bandwidth of the 13.5 GHz synthesizers.
Procedure
1.
Press UTILITY
, then
, then
, then
click Synthesizer Bandwidth Adj.
2.
Follow the instructions and prompts as they are displayed.
Source Adjustment
The source calibration is used to adjust your network analyzer for a flat source
power across its full frequency range. There are differences between each test
port; therefore, an adjustment is required for each port.
Equipment Used for the Source Adjustment
System
Service
Adjustment Routines
System
Service
Adjustment Routines
Equipment Type
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, 2.4 mm
8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
85056-60007
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...