58
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
8
Coordinating Multiple PXIe Oscilloscope Modules
C:\Program Files (x86)\Keysight\InfiniiVision>KtScopeCal.exe -r
Searching for connected modules...
Clock phase reset for module at slot 5...
Temperature change = 0.00
XML mean = 3081
Target = 3081
Iteration = 1
Mean = 3400
Mean change = 319
Iteration = 2
Mean = 2273
Mean change = -808
Iteration = 3
Mean = 3080
Mean change = -1
Clock phase reset completed.
Clock phase reset for module at slot 6...
Temperature change = 0.00
XML mean = 3117
Target = 3117
Iteration = 1
Mean = 4070
Mean change = 953
Iteration = 2
Mean = 3273
Mean change = 156
Iteration = 3
Mean = 2305
Mean change = -812
Iteration = 4
Mean = 1986
Mean change = -1131
Iteration = 5
Mean = 3109
Mean change = -8
Clock phase reset completed.
C:\Program Files (x86)\Keysight\InfiniiVision>
The red unlocked image will disappear.
4
Finally, to perform Sync mode trigger delay calibration on each trigger line and
save the calibration factors, run the
KtScopeCal -s
command. The options
are the same as when using the
KtScopeCal
command in Asynchronous mode.
For example:
C:\Program Files (x86)\Keysight\
InfiniiVision>KtScopeCal.exe -s PXI8::5::INSTR slot6 trig0
Trigger line PXI TRIG 0 bus segment configured: Isolated
Arm line PXI TRIG 1 bus segment configured: Isolated
Connect Aux Out from master module at slot 5 to channel 1 of master m
odule
and channel 1 of slave module at slot 6 using BNC tee adapter with
50ohm cables of equal length.