Coordinating Multiple PXIe Oscilloscope Modules
8
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
47
Slots 13-18:
Trig bus segment number 3
10-Slot Chassis
Slots 1-5:
Trig bus segment number 1
Slots 6-10:
Trig bus segment number 2
Sync Mode:
KtScopeCal -a
Align the reference clock signals for all modules and save factors.
Perform this action once for new or modified system.
KtScopeCal -p
Align the reference clock signals for all modules.
Perform this action each time InfiniiVision driver is connected.
KtScopeCal -r
Reset the reference clock signals for all modules.
Perform this action only when PXIe 100MHz reference clock is turned off.
This action is not required if -a or -p has been performed.
KtScopeCal -s PXI[a]::[b]::INSTR slot[c] trig[d] <bus[e]-[f]>
Calibrate trigger delay and save factors.
Perform this action after reference clock signals are aligned.
C:\Program Files (x86)\Keysight\InfiniiVision>
The chassis bus number can be found in the Keysight Connection Expert window:
1
Click the Keysight IO Libraries Suite icon in the task bar and choose
Connection
Expert
from the popup menu.
2
In the Keysight Connection Expert window,
PXI/AXIe Chassis > Chassis Content
tab, find the chassis bus number in the
Chassis Address
string:
Содержание InfiniiVision M9241A
Страница 1: ...Keysight M9241 42 43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide ...
Страница 6: ...6 Keysight M9241 42 43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide ...
Страница 10: ...10 Keysight M9241 42 43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide 1 Introduction ...