64
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
9
Applications Programming Interface (API) Overview
The oscilloscope's HiSLIP address can be cut-and-pasted from your remote
program or you can piece it together from the controller PC's IP address (or
hostname) and the oscilloscope's PXI address. For example if the controller
PC's IP address is "10.112.93.141" (or hostame is
"lab-pxi-3.cos.is.keysight.com") and the oscilloscope's PXI address is
"PXI37::0::0::INSTR", the oscilloscope's HiSLIP address is
"TCPIP0::10.112.93.141::hislip37-0.0::INSTR" (or
"TCPIP0::lab-pxi-3.cos.is.keysight.com::hislip37-0.0::INSTR").
If you are running the Interactive IO application on the controller PC, you can
use the "127.0.0.1" localhost IP address, for example,
"TCPIP0::127.0.0.1::hislip37-0.0::INSTR".
4
When connected, click
Device Clear
.