40
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
7
Make a Measurement
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the
trimmer capacitor on the probe for the flattest pulse possible.
On N2894A probes, the trimmer capacitor is located on the probe BNC
connector.
7
Connect probes to all other oscilloscope channels (that is, channel 2 of a
2-channel oscilloscope).
8
Repeat the procedure for each channel.
Turn On Measurements and Statistics
1
To turn on measurements, choose
Main Menu > Measure > Measurements
.
2
To turn on measurement statistics, choose
Main Menu > Measure > Statistics
.
You should see a display similar to this:
Perfectly compensated
Over compensated
Under compensated
NOTE
After a Default Setup, the Probe Comp signal is output to the Probe Comp terminal. However,
the oscilloscope can also output training signals on this terminal. See:
Main Menu > Training
Signals > Training Signals
. On InfiniiVision M924xA oscilloscope modules, the Probe Comp
terminal is also the Demo 1 terminal. There is no Demo 2 terminal (unlike some other
InfiniiVision X-Series oscilloscopes).