![Keysight InfiniiVision M9241A Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/keysight/infiniivision-m9241a/infiniivision-m9241a_startup-manual_1961660048.webp)
48
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
8
Coordinating Multiple PXIe Oscilloscope Modules
For each master module, slave module, and trigger line combination that will be
used:
1
Connect a MMCX-to-BNC cable from the master module Aux Out output to a
BNC tee adapter.
2
From the tee adapter, connect equal length cables to the master module
channel 1 input and the slave module channel 1 input.
3
Run the KtScopeCal.exe program.
For example:
C:\Program Files (x86)\Keysight\
InfiniiVision>KtScopeCal.exe PXI8::5::INSTR slot6 trig0
Trigger line PXI TRIG 0 bus segment configured: Isolated
Connect Aux Out from master module at slot 5 to channel 1 of master m
odule
Chassis Bus Number