Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
8
Coordinating Multiple PXIe Oscilloscope Modules
4
Finally, click
OK
.
Step 5: (Recommended) Phase-Lock Oscilloscope Sampling Clocks
It is also recommended that each module in the system phase-lock its timebase to
the PXIe 100 MHz reference.
For each module in the multiple-oscilloscope system:
1
Choose
Main Menu > Utilities > User Options Menu
.
2
In the User Options Menu, click the
Auxiliary I/O
softkey button.
CAUTION
The Keysight M9018A PXIe Chassis does not remember the trigger bus configuration
upon cycling power. After a power cycle, you must reconfigure the bus triggers by
using the M9018A Soft Front Panel (SFP) or by creating a persistent configuration by
using the AgM9018 API.