11 Using the SCPI Server
To connect the SCPI server to a specific E6610A instrument, type the command
CONTrol:INST:IPADdress xxx.xxx.xxx.xxx, where the x’s are replaced by the IP
address of the E6610A.
Note: The SCPI server does not display a prompt, and it does not return any
response unless a query was issued. To check for errors, use the command
SYSTem:ERRor? The SCPI server will display any error message or respond
with “No error.”
Instead of providing the IP address for the E6610A, you can also use the host name by
typing the command CONTrol:INST:HOST? KR12345678, where "KR12345678" is
replaced by the serial number of the desired E6610A. Note that this command must
be sent as a query with the question mark, and the SCPI server will reply with the host
name of the unit it has identified for connection.
To connect to the E6610A, send the command CONTrol:INST:CONN. You can confirm
the connection by sending the command *IDN? The response should include the
model number, serial number, and firmware revision of the connected instrument. You
are now ready to send SCPI commands to this instrument.
72
Keysight E6610A User and Programming Guide
Содержание E6610A
Страница 1: ...User and Programming Guide Keysight Remote Radio Head Tester E6610A...
Страница 8: ...8 Keysight E6610A User and Programming Guide...
Страница 10: ...10 Keysight E6610A User and Programming Guide...
Страница 12: ...12 Keysight E6610A User and Programming Guide...
Страница 20: ...20 Keysight E6610A User and Programming Guide...
Страница 36: ...36 Keysight E6610A User and Programming Guide...
Страница 64: ...64 Keysight E6610A User and Programming Guide...
Страница 67: ...Keysight E6610A User and Programming Guide 67...
Страница 108: ...This Page Intentionally Left Blank Keysight E6610A User and Programming Guide 108...