Tutorial
6
Output Characteristics
E3632A User’s Guide
155
Normal mode voltage noise is in the form of ripple related
to the line frequency plus some random noise. Both of these
are of very low value in the Keysight E3632A. Careful lead
layout and keeping the power supply circuitry away from
power devices and other noise sources will keep these values
low.
Common mode noise can be a problem for very sensitive
circuitry that is referenced to earth ground. When a circuit
is referenced to earth ground, a low level line — related AC
current will flow from the output terminals to earth ground.
Any impedance to earth ground will create a voltage drop
equal to the current flow multiplied by the impedance. To
minimize this effect, the output terminal can be grounded at
the output terminal. Alternatively, any impedances to earth
ground should have a complementary impedance to earth
ground to cancel any generated voltages. If the circuit is not
referenced to earth ground, common mode power line noise
is typically not a problem.
The output will also change due to changes in the load. As
the load increases the output current will cause a small drop
in the output voltage of the power supply due to the output
impedance R. Any resistance in the connecting wire will add
to this resistance and increase the voltage drop. Using the
largest possible hook- up wire will minimize the voltage drop.
Using the remote sense leads at the load will compensate for
lead resistance in the load leads.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...