Operation and Features
Programming the Overvoltage Protection
36
E3632A User’s Guide
3
Enter the OVP menu and set the desired trip level.
LEVEL 32.0 V
You will see the above message on the display when you
enter the OVP menu. Adjust the control knob for the
desired OVP trip level. Note that you cannot set the trip
levels to lower than 1.0 V.
4
Enable the OVP circuit.
OVP ON
You will see the above message after pressing
.
5
Exit the OVP menu.
CHANGED
The
CHANGED
message is highlighted for a second to show
that the new OVP trip level is now in effect. If the OVP
settings are not changed,
NO CHANGE
will be displayed.
The power supply will exit the OVP menu and the display
will return to the meter mode. Check that the
OVP
annunciator turns on.
Checking the OVP operation
To check OVP operation, raise the output voltage to near the
trip point. Then very gradually increase the output by
turning the knob until the OVP circuit trips. This will cause
the power supply output to drop to near zero, the
OVP
annunciator to flash, and the
CC
annunciator to turn on. The
OVP TRIPPED
message also appears on the display.
Clearing the overvoltage condition
When the OVP condition occurs (the
OVP TRIPPED
message is
shown on the display), the
OVP
annunciator flashes. When it
was caused by external voltage source such as a battery,
disconnect it first. The following steps show how to clear the
overvoltage conditions and get back to normal mode
operation. In the following steps, the display will go back to
OVP TRIPPED
if you let the display time- out after about
several seconds.
Over Voltage
Over Voltage
Over Voltage
Over Voltage
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...