3
Remote Interface Reference
Output Settings and Operation Commands
90
E3632A User’s Guide
Example
The following program segments show how to use the
VOLT
UP
or
VOLT DOWN
command to increase or decrease the
output voltage with the
VOLT:STEP
command.
VOLTage:TRIGgered
{<voltage>|MINimum|MAXimum}
This command programs the pending triggered voltage level.
The pending triggered voltage level is a stored value that is
transferred to the output terminals when a trigger occurs. A
pending triggered level is not affected by subsequent
VOLTage
commands.
VOLTage:TRIGgered? [MINimum|MAXimum]
This query returns the triggered voltage level presently
programmed. If no triggered level is programmed, the
VOLT
level is returned.
VOLT:TRIG? MAX
and
VOLT:TRIG? MIN
return the highest and lowest programmable triggered
voltage levels.
VOLTage:PROTection
{<voltage>|MINimum|MAXimum}
This command sets the voltage level at which the overvoltage
protection (OVP) circuit will trip. If the peak output voltage
exceeds the OVP level, then the power supply output is
shorted by an internal SCR. The Questionable Status register
“OV” bit is set (refer to
“SCPI status system” on page 108
An overvoltage condition can be cleared with the
VOLT:PROT:CLE
command after the condition that caused
the OVP trip is removed.
VOLT:STEP 0.01
Set the step size to 0.01 V.
VOLT UP
Increase the output voltage.
VOLT:STEP 0.02
Set the step size to 0.02 V.
VOLT DOWN
Decrease the output voltage.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...