2
Operation and Features
Constant Voltage Operation
28
E3632A User’s Guide
Constant Voltage Operation
To set up the power supply for constant voltage (CV)
operation, proceed as follows.
Front panel operation
1
Connect a load to the output terminals.
With power- off, connect a load to the (
+
) and (
–
) output
terminals.
2
Turn on the power supply.
The power supply will go into the power- on/reset state,
the output is disabled (the
OFF
annunciator turns on), the
15 V/7 A range is selected (the
15V
annunciator turns on),
and the knob is selected for voltage control.
To operate the power supply in the 30 V/4 A range, press
before proceeding to the next step. The
30V
annunciator turns on.
3
Set the display to the limit mode.
Notice that the
Limit
annunciator flashes, indicating that
the display is in the limit mode. When the display is in
the limit mode, you can see the voltage and current limit
values of the power supply.
Power
30V,4A
Display Limit
N O T E
In the constant voltage mode, the voltage values between the meter and
limit modes are the same, but the current values are not. Moreover, if the
display is in the meter mode, you cannot see the change of current limit
value when adjusting the knob. We recommend that you should set the
display to the limit mode to see the change of current limit value in the
constant voltage mode whenever adjusting the knob.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...