Remote Interface Reference
3
Output Settings and Operation Commands
E3632A User’s Guide
89
This command also increases or decreases the immediate
voltage level using the
UP
or
DOWN
parameter by a
predetermined amount. The command
VOLTage:STEP
sets
the amount of increment and decrement. Notice that a new
increment setting will cause an execution error - 222 (Data
out of range) when the maximum or the minimum rated
voltage is exceeded.
VOLTage? [MINimum|MAXimum]
This query returns the presently programmed voltage level of
the power supply.
VOLT? MAX
and
VOLT? MIN
return the
highest and lowest programmable voltage levels for the
selected range.
VOLTage:STEP {<numeric value>|DEFault}
This command sets the step size for voltage programming
with the
VOLT UP
and
VOLT DOWN
commands.
To set the step size to the minimum resolution, set the step
size to
DEFault
. The minimum resolution of the step size is
approximately 0.55 mV. The
VOLT:STEP? DEF
returns the
minimum resolution of your instrument. The immediate
voltage level increases or decreases by the value of the step
size. For example, the output voltage will increase or
decrease 10 mV if the step size is 0.01.
This command is useful when you program the power supply
to the allowed minimum resolution. At
*RST
, the step size is
the value of the minimum resolution.
VOLTage:STEP? [DEFault]
This query returns the value of the step size currently
specified. The returned parameter is a numeric value.
DEFault
gives the minimum resolution step size in unit of
volts.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...