2
Operation and Features
Programming the Overcurrent Protection
40
E3632A User’s Guide
Clearing the overcurrent condition
When the OCP condition occurs (the
OCP TRIPPED
message is
shown on the display), the
OCP
annunciator flashes. When it
was caused by external voltage source such as a battery,
disconnect it first. The following steps show how to clear the
overcurrent conditions and get back to normal mode
operation. In the following steps, the display will go back to
OCP TRIPPED
if you let the display time- out after about
several seconds.
1
Readjust the OCP trip level or the output current level.
Lower the output current level below the OCP trip point
after pressing
or raise the OCP trip level
by using the knob after pressing
.
2
Move to the clear mode.
OCP ON
You will see the above message after pressing
. If you changed the output voltage level,
press the
twice. Turn the knob to the right
until the
OCP CLEAR
message appears on the display.
3
Clear the overcoltage condition and exit this menu.
Now, when you press
again, the
DONE
message is displayed for a second and the
OCP
annunciator will not flash any more. The output will
return to meter mode.
Remote interface operation
or
Display Limit
Over Current
Display Limit
Over Current
Over Current
Over Current
Over Current
Over Current
Over Current
CURR:PROT {<current>|MIN|MAX}
Set the OCP level.
CURR:PROT:STAT {OFF|ON)
Disable or enable the
OCP circuit.
CURR:PROT:CLE
Clear the tripped
OCP circuit.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...