3
Remote Interface Reference
Output Settings and Operation Commands
86
E3632A User’s Guide
CURRent:STEP {<numeric value>|DEFault}
Set the step size for current programming with the
CURRent
UP
and
CURRent DOWN
commands. See the example as
shown in
.
To set the step size to the minimum resolution, set the step
size to
DEFault
. The minimum resolution of the step size is
approximately 0.12 mA. The
CURR:STEP? DEF
returns the
minimum resolution of your instrument. The immediate
current level increases or decreases by the value of the step
size. For example, the output current will increase or
decrease 10 mA if the step size is 0.01.
This command is useful when you program the power supply
to the allowed minimum resolution. At
*RST
, the step size is
the value of the minimum resolution.
CURRent:STEP? [DEFault]
This query returns the value of the step size currently
specified. The returned parameter is a numeric value.
DEFault
gives the minimum resolution of the step size in
unit of amps.
Example
The following program segments show how to use the
CURR
UP
or
CURR DOWN
command to increase or decrease the
output current with the
CURR:STEP
command.
CURR:STEP 0.01
Set the step size to 0.01 A.
CURR UP
Increase the output current.
CURR:STEP 0.02
Set the step size to 0.02 A.
CURR DOWN
Decrease the output current.
Содержание E3632A
Страница 1: ...Keysight E3632A DC Power Supply User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...