Calibration Procedures
Before beginning any adjustment procedures, the multimeter must be in
the “
UNSECURED
” state. To unsecure the multimeter, see “Calibration
Security Code” on page 73. Each adjustment should be followed by a
performance verification check for added confidence. We recommend
that you always adhere to the following general procedure.
•
Make sure you have read “Test Considerations” on page 64.
•
Unsecure the multimeter as outlined in “Calibration Security Code”
on page 73.
•
Perform the Zero Adjustment Procedure.
•
Perform the Gain Adjustment Procedure.
•
Perform the Verification Tests.
•
Secure the multimeter.
•
Note the new secure code and calibration count in your maintenance
records for future reference.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the
procedure has already been initiated. You can abort a calibration at
any time by pressing any front-panel key (except
Shift
). When
performing a calibration from the remote interface, you can abort a
calibration by issuing a remote interface device clear message or by
pressing the front-panel
LOCAL
key. Pressing the front/rear switch
during a calibration will also abort the calibration in progress.
Chapter 4 Calibration Procedures
Calibration Procedures
76
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...