7 Decrement the first digit until “0” is displayed.
You decrement or increment each digit independently. Neighboring
digits are not affected.
000.000,0 mVDC
8 Move the flashing cursor over to the “units” location.
Notice that the units are flashing on the right side of the display.
000.000,0 mVDC
9 Increase the displayed number by a factor of 10.
Notice that the position of the decimal point changes and the displayed
number increases by a factor of 10.
0.000,000 VDC
10 Save the change and turn off the menu.
The multimeter beeps and displays a message to show that the change
is now in effect. You are then exited from the menu.
The zero offset calibration procedure determines new calibration
constants for each function and range. Separate calibration is required
for the front and rear terminals.
<
<
∧
∨
Auto/Man
ENTER
3
Chapter 3 Menu Tutorial
A Front-Panel Menu Tutorial
57
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...