Resolution
(continued)
•
Front-Panel Operation: Select either the slow or fast mode for each
resolution setting. The default mode is 5 digits slow.
5: RESOLUTION (MEAS MENU)
See also “To Set the Resolution,” on page 37.
•
Remote Interface Operation: You can set the resolution using the
following commands.
Specify the resolution in the same units as the measurement
function, not in number of digits. For example, for dc volts, specify the
resolution in volts. For frequency, specify the resolution in hertz.
CONF:VOLT:DC 10,0.001
4
1
⁄
2
digits on the 10 Vdc range
MEAS:CURR:AC? 1,1E-6
6
1
⁄
2
digits on the 1 A range
CONF:FREQ 1 KHZ,0.1 Hz
1000 Hz input, 0.1 Hz resolution
VOLT:AC:RES 0.05
50 mV resolution on the ac function
CONFigure:<
function
> {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
MEASure:<
function
>? {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
<
function
>:RESolution {<
resolution
>|MIN|MAX}
Chapter 4 Calibration Procedures
Understanding Resolution
88
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...