![Keysight 34401A Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/keysight/34401a/34401a_service-manual_1961480062.webp)
Menu Example 2
Some commands in the menu require that you enter a numeric parameter
value. The following steps show you how to enter a number in the menu.
In this example, you will set the calibration value to 0.0 volts. For this
example you must apply a short between
HI-LO Sense
and
HI-LO Input
.
Caution
Completing this example will perform a zero calibration. Refer
to chapter 4, “Calibration Procedures,” before attempting this example.
1 Turn on the menu.
You enter the menu on the “menus” level. The
MEAS MENU
is your first
choice on this level.
A: MEAS MENU
2 Move across to the
CAL MENU
choice on this level.
There are six menu group choices available on the “menus” level. Each
choice has a letter prefix for easy identification (
A:
,
B:
, etc.).
F: CAL MENU
<
Shift <
On/Off
3
Chapter 3 Menu Tutorial
A Front-Panel Menu Tutorial
55
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...