Optional AC Performance Verification Tests
These tests are not intended to be performed with every calibration.
They are provided as an aid for verifying additional instrument
specifications. There are no adjustments for these tests; they are
provided for performance verification only.
Configuration: AC volts
6
1
⁄
2
digit
AC FILTER
slow (
MEAS MENU
)
1
Make sure you have read “Test Considerations” on page 64.
2
Select each function and range in the order shown below.
Compare measurement results to the appropriate test limits shown
in the table. (Be certain to allow for appropriate source settling.)
Input
(Front)
Input
Frequency
34401A
Range
Error From Nominal
24 hour
90 day
1 year
1 V
1 V
1 V
1 V
1 V
1 V
10 V
1 V
100 mV
20 Hz
1 kHz
20 kHz
50 kHz
100 kHz
300 kHz
1 kHz
1 kHz
1 kHz
1 V
1 V
1 V
1 V
1 V
1 V
10 V
10 V
10 V
±
600
µ
V
±
600
µ
V
±
600
µ
V
±
1.4 mV
±
6.3 mV
±
45 mV
±
6 mV
±
2.4 mV
±
12 mV
±
800
µ
V
±
800
µ
V
±
800
µ
V
±
1.6 mV
±
6.8 mV
±
45 mV
±
8 mV
±
3.5 mV
±
13 mV
±
900
µ
V
±
900
µ
V
±
900
µ
V
±
1.7 mV
±
6.8 mV
±
45 mV
±
9 mV
±
3.6 mV
±
13 mV
Chapter 4 Calibration Procedures
Optional AC Performance Verification Tests
72
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...