Self-test procedures are numbered in the order in which they are
executed. The order is designed to build confidence from a small kernel
towards progressively more complex internal configurations.
Many individual self-test procedures may be executed from the front
panel by selecting the desired test number from the parameter list of
the
TEST
command (in the
SYS MENU
).
Note
K101 is always open during self-test so that signals present on the
input terminals are disconnected. However, ac signals may still feed
through K101 causing self-test failures as mentioned earlier in this section.
601
Front panel does not respond
The main
CPU U500
attempts to
establish serial communications with the front panel processor
U600
.
During this test,
U600
turns on all display segments. Communication
must function in both directions for this test to pass.
602
RAM read/write failed
This test writes and reads a 55
H
and AA
H
checker board pattern to each address of ram
U503
. Any incorrect
readback will cause a test failure. This error is only readable from the
remote interface.
603
A/D sync stuck
The main
CPU
issues an A/D sync pulse to
U500-38
and to
U501
where
ADC
slope counters are latched. A failure is detected
when a sync interrupt is not recognized and a subsequent time-out occurs.
604
A/D slope convergence failed
The dc section is configured to the
measure zero state (
MZ
) in the 10 V range. This test checks whether the
ADC
integrator produces nominally the same number of positive and
negative slope decisions (
±
10%) during a 20 ms interval.
605
Cannot calibrate rundown gain
This test checks the nominal gain
between the integrating ADC and the U500 onchip ADC. This error is
reported if the procedure can not run to completion due to a hardware failure.
606
Rundown gain out of range
This test checks the nominal gain
between the integrating
ADC
and the
U500
onchip
ADC
. The nominal
gain is checked to
±
10% tolerance.
6
Chapter 6 Service
Self-Test Procedures
121
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...