Pass/Fail Output
(continued)
Installation Procedure
The assembly drawings and schematics are located in chapter 9,
“Schematics.”
1
Unscrew the two rear bezel captive screws (see the mechanical
disassembly drawing on page 9-3).
2
Remove one screw from bottom cover (see the mechanical
disassembly drawing on page 9-3).
3
Slide the metal cover off of the chassis (see the mechanical
disassembly drawing on page 9-3).
4
Install
JM710
and
JM711
located adjacent to the power supply
connector,
P751
. A bare wire may be used if a 0 ohm surface
mount resistor is not available. (See the component locator
diagram for the 34401-66501 Main PC Board on page 9-5.)
5
Reassemble the multimeter. Make sure to reinstall the bottom
cover screw.
Do not use the RS-232 interface if you have configured the multimeter
to output pass/fail signals on pins 1 and 9. Internal components on the
interface circuitry may be damaged.
Chapter 6 Service
To Connect the Pass/Fail Output Signals
116
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...