To Make dB Measurements
Each dB measurement is the difference between the input signal and a
stored relative value, with both values converted to dBm.
dB
= reading in dBm
–
relative value in dBm
•
Select
DC V
or
AC V
.
•
The first reading taken after you enable dB measurements is
converted to dBm and is stored as the relative value in the
dB Relative Register. Any previously stored value is replaced
with the new value.
•
After enabling dB operations, you can edit the relative value by
pressing
Shift >
(Menu Recall). This takes you to the “
dB REL
”
command in the
MATH MENU
(only if dB is enabled). Go down to
the “parameter” level, and then edit the value displayed.
•
The register is cleared when you change functions, turn dB off,
turn off the power, or perform a remote interface reset.
Enables dB operation;
Press again to disable.
Math
annunciator is on when
dB operation is enabled.
To read / edit the dB relative value, use the MATH menu.
Chapter 2 Quick Start
To Make dB Measurements
40
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...