Caution
Zero offset calibration using a multifunction calibrator is
NOT
recommended. The calibrator and cabling offset can be large and unstable
causing poor offset calibration of the Agilent 34401A or any multimeter.
Note
These offset tests should be verified for both the front and the rear
input terminals.
Input
(Front/Rear)
Function
Quick
Check
34401A
Range
Error From Nominal
24 hour
90 day
1 year
Open
Open
Open
Open
Short
Short
Short
Short
Short
Short
Short
Short
Short
Short
Short
Short
DC Current
DC Volts
2-Wire
Ohms[1]
and
4-Wire
Ohms
Q
Q
Q
10 mA
100 mA
1 A
3 A
100 mV
1 V
10 V
100 V
1000 V
100
Ω
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
10 M
Ω
100 M
Ω
±
1
µ
A
±
4
µ
A
±
60
µ
A
±
600
µ
A
±
3
µ
V
±
6
µ
V
±
40
µ
V
±
600
µ
V
±
6 mV
±
3 m
Ω
±
5 m
Ω
±
50 m
Ω
±
500 m
Ω
±
10
Ω
±
100
Ω
±
10 k
Ω
±
2
µ
A
±
5
µ
A
±
100
µ
A
±
600
µ
A
±
3.5
µ
V
±
7
µ
V
±
50
µ
V
±
600
µ
V
±
10 mV
±
4 m
Ω
±
10 m
Ω
±
100 m
Ω
±
1
Ω
±
10
Ω
±
100
Ω
±
10 k
Ω
±
2
µ
A
±
5
µ
A
±
100
µ
A
±
600
µ
A
±
3.5
µ
V
±
7
µ
V
±
50
µ
V
±
600
µ
V
±
10 mV
±
4 m
Ω
±
10 m
Ω
±
100 m
Ω
±
1
Ω
±
10
Ω
±
100
Ω
±
10 k
Ω
[1] For 2-wire ohms, these errors assume you are using Math Null. Without Math Null,
an additional 200 m
Ω
of error must be added.
Q
: Quick performance verification test points.
Chapter 4 Calibration Procedures
Zero Offset Verification
68
Содержание 34401A
Страница 1: ...Keysight 34401A 6 Digit Multimeter Service Guide ...
Страница 2: ......
Страница 17: ...Contents 10 ...
Страница 18: ...1 Specifications 1 ...
Страница 33: ...26 ...
Страница 34: ...2 Quick Start 2 ...
Страница 54: ...3 Menu Tutorial 3 ...
Страница 65: ...58 ...
Страница 66: ...4 Calibration Procedures 4 ...
Страница 98: ...5 Theory of Operation 5 ...
Страница 115: ...108 ...
Страница 116: ...6 Service 6 ...
Страница 132: ...7 Replaceable Parts 7 ...
Страница 143: ...136 ...
Страница 144: ...8 Backdating 8 ...
Страница 147: ...34401 66522 Display Assembly Component Locator Chapter 8 Backdating Agilent 34401A Multimeter 140 ...
Страница 149: ...9 Schematics 9 ...
Страница 165: ...This information is subject to change without notice Keysight Technologies 1991 2014 Edition 9 August 2014 34401 90013 ...