U3042AM08
Service Information
RF Switching Failures
Typical failures may be the loss of RF switch control or a non RF failure as outlined in
the prior section. The internal RF switches used in this instrument are controlled by
the +21 Vdc from the Switch Driver board.
1.
Confirm that the analyzer and test set are properly connected using the Test Set
I/O cable (N4011-21002). See
.
2.
Using Multiport Mode (Option 551), verify that the analyzer’s firmware is correct,
and that the test set file is loaded into the analyzer. Refer to
3.
Using the I/O command values, confirm that the correct address and data values
are used. Refer to
“Address and Data Values” on page 5
4.
Follow the
procedures (if not already done so).
5.
Front Panel R and S indicator LED Check.
a.
Confirm that the test set Controller board Status LEDs are On. Refer to
. If the LEDs are Off remove the Switch Driver board, if the LEDs
are remain Off replace the Controller board.
b.
If the Status LEDs are On and the front panel Active LED is On, suspect the
front panel LED board or the ribbon cable (replace as needed).
86
Keysight U3042AM08 User's and Service Guide